A lampholder gauge is a small tool with several working surfaces, and each one carries a different load in use. They wear at different rates, and the direction of wear differs. Inspecting the tool as a single object hides that.
| Surface | Function | Wear direction |
|---|---|---|
| thread section | checks cap thread | pitch diameter grows |
| datum face | sets seating depth | face becomes dished or rounded |
| contact pin | checks centre contact position | pin shortens or rounds |
This is the surface everyone checks, because it is the one that visibly wears. Each cap screwed on polishes the flanks, the pitch diameter grows, and the gauge becomes more permissive.
The rate depends on use and on cleanliness. A gauge used on caps with flash or mould release will wear faster than one used on clean caps, because grit between the flanks acts as an abrasive.
The face that the cap seats against is checked far less often, and it matters as much. If the face becomes dished, the cap seats deeper than it should and the gauge reports a pass for a cap with a short thread. If a burr forms on the rim, the cap seats shallow and a good cap is rejected.
The face is also the surface most likely to be damaged by handling: a gauge dropped on a bench lands on its face more often than on its thread.
Checking it is simple: a straightedge across the face shows dishing as a gap in the middle, a burr as a rocking point.
On gauges that include a centre contact check, the pin that represents the contact position wears at its tip. A pin that has shortened by a few hundredths lets a cap pass whose contact sits too deep.
This surface is the least inspected of the three, because it is small and often hidden inside the gauge body. It is also the one whose wear produces the least visible symptom — the cap still screws on, still seats, and the failure only appears when a lamp does not make contact.
The pattern tells you:
If caps increasingly pass when they should fail, and the thread looks polished, the thread section has grown.
If caps that seat properly are rejected, and the datum face shows a burr, the face is the cause.
If lamps intermittently fail to light but the caps gauge normally, suspect the contact pin.
The thread section should be checked on a schedule tied to use. The datum face should be checked whenever the gauge is dropped or handled roughly — which is not a schedule, it is an event, and it needs to be written into the work instruction.
The contact pin should be checked whenever the gauge is sent for calibration, and additionally whenever a lamp-contact complaint is investigated.
For each surface, a measurement rather than a verdict: pitch diameter at a defined point, face flatness (feeler gauge result), pin protrusion. Three numbers, recorded at each inspection, give a trend that tells you which surface is moving and how fast.
A record that says "gauge OK" for all three surfaces has recorded nothing.