A lamp cap gauge is defined by a sheet, not by a catalogue number. The sheet carries a figure, dimensions with tolerances, and a revision suffix. When the sheet is revised, the gauge in the gauge room does not change, its calibration certificate does not change, and nothing on the shop floor looks different — but the question it answers may have.
A designation such as 7006-27D-3 has three parts, and each carries information an inspector needs:
The distinction between a letter change and a number change is the part that causes mistakes. 27D and 27E are two different gauges; 27D-2 and 27D-3 are the same gauge at two revisions of the sheet. Buying "the latest 27D" gets the current revision; buying "27E" because it looks similar gets the wrong instrument.
The sheet number itself is not a serial. It is allocated in blocks that follow cap families, and the family is stated on the sheet: a 7006-27 series gauge checks an E26 cap whose thread is nominally 26 mm across, and a 7006-10 series gauge covers a B22 bayonet cap whose shell is 22 mm. The correspondence is not arithmetic — the E14 gauges sit in a different block — so the number has to be read from the sheet, not guessed from the thread size.
Revisions are not editorial. In practice they do one of four things:
Only the third changes what an existing gauge measures. The first two change what a compliant gauge should be, and the fourth means the old gauge cannot check the new part at all. Put a number on it: on a cap whose thread limits are held to a few hundredths of a millimetre, a revision that moves a limit by 0.02 mm is the difference between a cap that passes and one that does not, and the gauge that measured it last month still reports the old answer.
"Show me the sheet this gauge was made to." A gauge with a certificate that says "in accordance with IEC 60061-3" and no revision number cannot answer it. Three records close the gap:
When they do not match, the decision is not automatic. A gauge to revision 2 checking a cap that has not changed since revision 1 is still valid; a gauge to revision 2 checking a cap defined at revision 3 is not.
Most gauge rooms end up with a mixture of revisions, because gauges outlive standards. A workable policy has four parts:
| Classify | Mark each gauge with its sheet revision, physically, not only in a database |
| Map | Record which cap variants each revision covers, so a superseded gauge is used only where it is still valid |
| Review | Check the sheet revision at every calibration, and again whenever a new cap variant is introduced |
| Retire | Withdraw a superseded gauge when the parts it covers are no longer produced, and record the date |
Calibration answers a narrow question: does this gauge still match the dimensions on the sheet it was made to? It does not answer whether that sheet is the right one. A gauge can be perfectly calibrated to a superseded sheet and still be the wrong instrument for the parts now being produced.
This is the same logic that applies to any limit gauge, and it is worth stating in the procedure: calibration confirms the instrument, not the specification.
The lamp cap gauge range lists gauges by 7006 designation and sheet revision; the catalogue home indexes the lampholder gauges used alongside them.