Cap Gauge Uncertainty: A Three Micrometre Budget at 20 °C

Home News

Cap Gauge Uncertainty: A Three Micrometre Budget at 20 °C

Good quality IEC/EN 60065 Standard for sales
Good quality IEC/EN 60065 Standard for sales
Our goods has safely arrived at our residence. Thank you for your excellent services. I hope we can cooperate again in the future.

—— Wira

I'm Online Chat Now
Company News
Cap Gauge Uncertainty: A Three Micrometre Budget at 20 °C

Every gauge certificate carries a measurement result and, if it is worth the paper, an uncertainty. In cap inspection the uncertainty is often larger than the tolerance it is being compared with, and the reason is almost always temperature rather than the instrument.

Building the budget takes twenty minutes and tells you which of five contributors to attack.

The five contributors

ContributorTypical magnitudeHow to reduce it
Reference standard uncertainty1 µm to 2 µmA better setting rod or master
Repeatability of the measurement1 µm on a good micrometerSame operator, same technique, three readings
Temperature difference from 20 °C11 µm per metre per K for steelSoak the part, measure fast
Thermal expansion mismatchDepends on materialsMeasure steel against steel
Geometry and elastic deformation1 µm to 3 µm on a curved surfaceUse the right anvil, control the force

Steel expands about 11 micrometres per metre per kelvin, which sounds small until it is applied to a 100 mm setting rod: 5 K away from 20 °C is 5.5 µm, which is the entire gauge tolerance on many cap dimensions. The part and the gauge both move, but not necessarily by the same amount, because the part may be brass and the gauge steel.

Why brass and steel disagree

Brass expands at about 19 micrometres per metre per kelvin, nearly twice steel. A brass cap measured at 26 °C is bigger than the same cap at 20 °C by more than the go gauge allowance on a nominal 26 mm diameter. In summer, in an inspection room without temperature control, a batch can fail in the morning and pass in the afternoon without anything changing in the process.

The practical rule is to soak the parts in the inspection room for long enough that a probe on the surface reads the room temperature, not to measure them as they arrive from the moulding shop.

Putting the numbers together

  • Write each contributor as a standard uncertainty, not as a limit.
  • Combine them as the root sum of squares, which is the standard approach and is easy to do in a spreadsheet.
  • Multiply by the coverage factor of 2 for a 95 percent interval.
  • Compare the result with the gauge tolerance. If it is more than a quarter, the measurement is not fit for the decision.

What to do when the budget is too large

Three routes, in order of cost. Control the temperature, which is a room and a soak period. Reduce the number of transfers by using a comparator against a master of the same material. Or change the decision rule: gauge to a guard band so that parts within the uncertainty of the limit are treated as conditional rather than accepted.

None of these requires a better micrometer. Most cap gauging problems are thermal, and the instrument is usually the smallest term in the budget. A cap gauge set with a certificate stating both the uncertainty and the reference temperature is the starting point, and the site index lists the setting masters by cap family.

Standards referenced: ISO/IEC Guide 98-3; ISO 14253-1.